测量范围 |
剂量当量 AT3509 AT3509A:Hp(10)..................1 ?Sv - 10 Sv AT3509B:Hp(10),Hp(0.07)..............1 ?Sv - 10 Sv AT3509C:Hp(10),Hp(0.07)..............1 ?Sv - 10 Sv |
剂量率当量 AT3509 AT3509A:Hp(10)...........0.1 ?Sv/h - 1 Sv/h AT3509B:Hp(10),Hp(0.07)........0.1 ?Sv/h - 1 Sv/h AT3509C:Hp(10),Hp(0.07)........0.1 ?Sv/h - 5 Sv/h 剂量测量误差(无伴随β辐射)...................... ±15% 137Cs校准误差.............................................. ±5 % |
剂量率测量误差 0.1 to 1 ?Sv/h.............................................. ±25 % 超过1 ?Sv/h.................................................. ±15% |
能量响应 AT3509 AT3509B AT3509C ......................15 keV - 10 MeV AT3509A .....................................30 keV - 10 MeV |
能量响应差异 Hp(10) 15 keV — 1.5 MeV.................................... ±25 % 1.5 MeV — 10 MeV.................................... ±60 % Hp(0.07) AT3509B 15 keV — 300 KeV.....................................±25 % |
±75°角范围内各向异性 137Cs .......................... ............................... ±20 % 241Am .......................................................... ±50 % |
报警限值 剂量:30 ?Sv;200 ?Sv;1 mSv;4.2 mSv;12.5 mSv; 50 mSv;100 mSv;1 Sv 剂量率:0.3 Sv/h;3 ?Sv/h;30 ?Sv/h;300 ?Sv/h; 3 mSv/h;30 mSv/h;300 mSv/h;1 Sv/h 也可自定义其它限值 |
响应时间 Hp(10) > 10 ?Sv/h.............................................. 5 s 辐射超负荷......................................zui高达10 Sv/h |
操作环境 温度范围....................... ..................... -10°,+40°C 相对湿度(+ 40°C)......................................... 90 % 防震(从1.5 m处落下) 防护等级.................... ..................................... IP54 |
响应时间 Hp(10) > 10 ?Sv/h...................................... 5 s |
电源要求 2 cells type AAA (LR03)或蓄电池 连续操作时间 Hp(10) < 100 ?Sv/h....................................... 300 h 尺寸............................................ 104×58×23 mm 重量(不带电池)............................................ 100 g |